Summary: The following are Cleaning, Sterilizing, and maintaining surgical instruments guidelines.
The use of distilled or deionized water and a neutral pH cleaning solution is recommended for all these procedures.

1. Rinsing- Immediately after surgery, rinse instruments under warm (not hot) water. It may be helpful to use a nylon toothbrush to rinse the lock boxes and joints of the instrument. Be sure to remove all blood, body fluids, and tissue.

2. Cleaning- If you do not clean your instruments immediately after rinsing, instruments should be submerged in a solution of water and neutral pH (7) detergent. They should never be placed in saline solution, as it may cause corrosion and eventually irreversible damage to the instrument.

A. Ultrasonic Cleaning-
We recommend you manually clean micro & delicate instruments.
Instruments should be processed in a cleaner for the full recommended cycle time- usually 5-10 minutes.
Place instruments in an open position in the ultrasonic cleaner. Make sure that sharp blades such as scissors, knives, osteotomes, etc. do not touch other instruments.
All instruments have to be fully submerged.
Do not place dissimilar metals (stainless, copper, chrome plated, etc.) in the same cleaning cycle.
Change solution frequently, at least as often as recommended by the manufacturer.
Rinse instruments after ultrasonic cleaning with distilled or deionized water to remove the ultrasonic cleaning solution.

B. Automatic Washer Sterilizers-
Follow manufacturers’ recommendations. Make sure instruments are lubricated after the last rinse cycle and before the sterilization cycle.

C. Manual Cleaning-
Most instrument manufacturers recommend ultrasonic cleaning as the best and most effective way to clean surgical instruments, particularly those with hinges, locks, and other moving parts. If ultra-sonic cleaning is not available observe the following steps.
I. Use stiff plastic cleaning brushes (nylon etc.).
Do not use steel wool or wire brushes except specially recommended stainless steel wire brushes for instruments such as bone files, or on stained areas in knurled handles.
II. Use only neutral pH (7) detergents such as the Roboz Instrument Cleaner (IC-1000) or the Roboz Enzymatic Cleaner (EC-1000.) If not rinsed off properly after cleaning, low pH detergents will break the stainless protective surface and cause black staining.
High pH detergent will cause a surface deposit of brown stain (this deposit may look like rust) which will also interfere with the smooth operation of the instrument.
III. Brush delicate instruments carefully and, if possible, handle them separately from general instruments.
IV. Make sure all instrument surfaces are visibly clean and free from stains and tissue.
This is a good time to inspect each instrument for proper function and condition.
Check the following:
Scissor blades glide smoothly from open to closed (they must not be loose when in a closed position). Test scissors by cutting them into thin gauze. Three-quarters of the length of the blade should cut to the scissor tips, and not hang up.
Forceps should have properly aligned tips.
Hemostats and needle holders should not show any light between the jaws. They should lock and unlock easily and the joints should not be too loose. Check needle holders for wear at the tips.
Suction tubes should be clean inside.
Retractors should open, close, and lock properly.
Blades of all cutting edges should be sharp and undamaged.
V. After manually scrubbing instruments, rinse them thoroughly under running water (distilled water is best). While rinsing, open and close scissors, hemostats, needle holders, and other hinged instruments to make sure the hinge areas are rinsed out, as well as the outside of the instruments.

3. After Cleaning- If the instruments are to be stored, let them air dry and store them in a clean and dry environment.

4. Autoclaving- If instruments are to be reused or autoclaved:
A. Lubricate all instruments which have any metal-to-metal action such as scissors, hemostats, needle holders, and retractors.
B. Use disposable paper or plastic pouches to sterilize individual instruments. Ensure you use a wide enough pouch (4” or wider) for instruments with ratchet locks such as needle holders and hemostats so the instrument can be sterilized in the open and unlocked position.
Suppose you are autoclaving instrument sets unlock all instruments and sterilize them in an open position. Place heavy instruments at the bottom of the set (when two layers are required).
Never lock an instrument during autoclaving. It will not be sterile as the steam cannot reach the metal-to-metal surfaces. The instrument might develop cracks in hinged areas caused by the heat expansion during the autoclave cycle.
Do not overload the autoclave chamber. Pockets may form that do not permit steam penetration. Place a towel on the bottom of the pan to absorb excess moisture during autoclaving.